<6. Operation and Parameters>

6-11

IM 11M10B01-01E

calibration. However, there are the following limits to the O

2

concentration. Be sure to perform

calibration in that range of the O

2

concentration.

O

2

concentration limits: 0.9 ppm or more and 12% or less, or between 35% and 100%.

4) Air calibration “AIR”

This is air-only calibration. Ues the cleanest air possible to perform calibration.

NOTE: If you press the [MODE] key during calibration to exit the calibration mode, that calibration

becomes invalid.

If an error occurs during calibration, the ERR/ALM lamp turns on, and an error/alarm code is displayed

on the sub-display. That calibration may become invalid depending on the error factor.
Figure 6.14 shows the calibration method (including the calibration gas concentration setting). With

respect to the input method of a numerical value, decimal point, and unit, see Section 6.6.

C AL

AIIR

CAL

[ENT]

ALL

CAL

[ENT]

AIR

PPM

[ENT]

20.60

OK

PPM

[ENT]

1

0.38

10.00

PPM

[ENT]

10.38

PPM

[ENT]

[ENT]

0

989

1000

PPM

[ENT]

1001

989

PPM

AIR

PPM

[ENT]

AIR

PPM

[ENT]

[ENT]

1Pnt

CAL

[ENT]

5

0.25

1Pnt

PPM

[ENT]

50.25

PPM

[ENT]

50.25

OK

PPM

[ENT]

2Pnt

CAL

[ENT]

1

0.00

ZEro

PPM

[ENT]

10.24

PPM

[ENT]

9

0.25

SPAn

PPM

[ENT]

90.18

90.25

PPM

[ENT]

90.25

OK

PPM

[ENT]

M O D E

OK

[ENT]

PPM

50.07

10.00

20.78

10.21

20.60

20.95

20.60

%

%

%

%

%

%

%

%

%

%

%

%

%

%

%

%

%

Hold down

for 2 seconds.

If an error occurs, and “ERR” appears in the sub-display,

calibration may be not accepted.

Execute 1-point calibration

Execute 2-point calibration

Execute Air calibration

Execute 3-point (All) calibration

To “ALL”

Enter the gas concentration

at any calibration point.

Enter the numerical value,

decimal point, and unit.

Enter the numerical value,

decimal point, and unit.

Enter the numerical value,

decimal point, and unit.

Wait for stabilization after

sampling calibration gas.

Display the gas

concentration at the calibration point.

Calibration successful

Return to “CAL.”

Enter the gas concentration at zero point.

Wait for stabilization after

sampling zero gas.

Display the gas

concentration at zero point.

Enter the gas concentration at span point.

Wait for stabilization

after sampling span gas.

Display the gas

concentration at span point.

Calibration successful.

Return to “CAL.”

Wait for stabilization

after sampling air.

Calibration successful

Return to “CAL”

Enter the 10-ppm

gas concentration.

Enter the 1000-ppm

gas concentration.

Enter the numerical value only.

Enter the numerical value only.

Wait for stabilization

after sampling

10-ppm gas.

Wait for stabilization

after sampling

1000-ppm gas.

Display the

10-ppm gas concentration.

Display the

1000-ppm gas

concentration.

Return to “CAL.”

Check the air concentration

(setting is not needed).

Wait for stabilization

after sampling air.

Calibration successful.

(Note) If an error occurs during calibration, that calibration may become invalid. With respect to the action

in the event of an error, see Section 7.2.

To “1Pnt”

Figure 6.14

Calibration “CAL”