background image

<

4. Piping and Wiring>

4-3

IM 11M10B01-01E

MS1

MS2

MS3

MSI1

MSI2

MSICOM

MSCOM

1st flow path

2nd flow path

3rd flow path

OX400 Rear Terminal (M3 screw)

The following terminals are added for the multi-selector “/MS” option.

The customer needs to supply a switching device and carry out the necessary wiring.

Contact output
for switching of
measurement flow
path

Contact output
for measurement
flow path data

Common

Common

Common

PUMP

OFF

mA

G

V

RC2
RC3
RCCOM

RC1

FAIL

DO

Pump ON/OFF

Contact input

(available when suffix code of

with built-in pump “-P” specified)

Secondry output
(0-1/0-5/0-10 V DC)

Power cord (two-pole plug with earthing contact)

Primary output
(4 to 20 mA DC)

+

-

+

-

Alarm

(Oxygen concentration high/low)

Contact output

Error (Fail)

Contact output

Range marker

Contact output

(Note 2) (Note 3)

(Note 1)

Note 1: Use the earthing contact of power cord to ground to earth.

Use the supplied power cord only.

Note 2: Ground the measurement output signal line shield on the receiving side.

The G-terminal is connected to a ground pin. Use this if the line shield cannot be grounded on the receiving side.

Be very careful not to ground the line at two points.

Note 3: The signal output and contact input cable must be no longer than 30 m for CE marking. RS232 cable must be no longer than 3 m.

(NOT using)

Figure 4.3

External Wiring Diagram

4.2.2 Signal Wiring

Table 4.1 shows the signal assignments of terminals on the rear panel. Perform wiring as necessary.
It is recommended to use an insulation sleeve crimp terminal (for M3 screws) for cable termination.

Table 4.1

List of Input and Output Signals

Type of Signal

Terminal Marking

Description

Screw

Wiring Limit

Primary output terminal

mA + -

4-20 mA current output

M3

(Note)

Secondary output terminal

V + -

0-1, 0-5, 0-10 V DC voltage output

M3

(Note)

Range marker terminal

RC1, RC2, RC3,

RCCOM

For auto range recognition

M3

Contact input terminal

PUMP OFF

For pump remote ON/OFF

M3

(Note)

Error contact output terminal

FAIL

Output in the event of an error (ERR)

M3

Alarm contact output terminal

DO

Output in the event of an oxygen

concentration alarm (ALM7)

M3

Contact output terminal for

multi-selector

MS1, MS2, MS3,

MSCOM

MSI1, MSI2,

MSICOM

For measurement flow path switching

(1 to 3)

For information on flow path under

measurement

M3

(Note) The current or voltage signal output cable length must be no longer than 30 m for CE use.