<7. Inspection and Maintenance>

7-6

IM 11M10B01-01E

4) Insert the ring

1

and holding plate

4

in this order into the sensor. Hold the sensor with

fingers and rotate the joint

2

right and left to push it into the sensor slowly.

Lead wire connector

3

Heater

Sensor

1

4

2

5) Rotate and tighten the ring

1

into the joint

2

. Secure the holding plate

4

with two screws

(7).

Connect the lead wire connector of the sensor to the PCB board.

Lead wire connector

Heater

Sensor

7

Sensor constant tag

3

1

4

2

6) When the installation of the sensor is finished, take a note of the sensor constant on the sensor

constant tag attached to the sensor lead wire, and then close the cover.

7) Turn on the power, and when “HEAt” appears and countdown starts, enter sensor constant

“SEnS.” As for how to enter it, see 6.1.7 “Setting Sensor Constant.”

Be sure to perform calibration after replacing the sensor.

7.4 How to Replace Activated Carbon Filter

Activated charcoal filter will be provided when “/A” option is specified. Filter should be installed

between gas sampling line and gas inlet of OX400. Filter has a life time and need to replace

periodically, its replacement interval vary depending on the application. As for the part number, see

CMPL at the end of this manual.