7. Inspection and Maintenance>

7-5

IM 11M10B01-01E

7.3.2 Installing Sensor

1) Pass a new sensor through the heater.

Sensor

Lead wire connector

Heater

2) Insert the nut

3

, holder

5

, and O-ring

6

in this order from the front end of the sensor. At

this point, set the O-ring in the position about 3 mm from the sensor front end (It is recommended

to replace the O-ring with a new one).

O-ring

Lead wire connector

Heater

5

3

6

Approx. 3 mm

Sensor

Marking

13 mm

3) Insert the sensor into the connector and tighten the nut

3

firmly with fingers. At this point, make

sure that the lead wire of the sensor is located in the upper right position.

Lead wire connector

Sensor

Heter

Move the sensor back and forth to align

the nut with the sensor marking line.

3

Figure 7.3

Installing Sensor