1 inspection in the event of an alarm (alm), 2 inspection in the event of an error (err) – Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 55

Advertising
background image

<

7. Inspection and Maintenance>

7-2

IM 11M10B01-01E

7.2.1 Inspection in the Event of an Alarm (ALM)

If an error is detected by the self-diagnosis function of the OX400 during operation, the ERR/ALM

lamp flashes and an alarm (ALM) code is displayed on the sub-display. If an error is detected, take

action according to the Table 7.1.

Table 7.1

Alarm (ALM) Conditions and Actions

Display

Content

Diagnosis Condition

Action

ALM1

Unstable heater

temperature

The heater temperature changed by ±5°C or

more from the standard temperature.

Check to make sure the

sensor gas flow rate or

power supply is not unstable

ALM2

Sensor error

The sensor electromotive force during

calibration became greater than ±30% relative

to the standard.

Check the calibration gas O

2

concentration value

Re-calibrate

Replace the sensor

ALM3

Electromotive force error

A sensor electromotive force error of ±15 to

±30% relative to the standard occurred during

calibration.

Check the calibration gas O

2

concentration value

Re-calibrate

Replace the sensor

ALM4

Asymmetric voltage error

The air electromotive force became greater than

±10 mV during calibration.

Calibration up to ±30 mV is possible.

Calibration beyond ±30 mV is not possible.

Check the air concentration

Re-calibrate

Replace the sensor

ALM5

Calibration error

The calibration coefficient became greater than

the specified values (zero: ±50, span: 1 ±0.2).

When zero is ±50 to ±75 and span is 1±0.2 to

1±0.4, calibration is possible

When zero exceeds ±75 and span exceeds 1

±0.4, calibration is not possible

Check the calibration gas O

2

concentration

Re-calibrate

Replace the sensor

ALM6

Sensor resistance value

error

The sensor impedance (cell resistance value)

became greater than 1 kΩ during calibration.

Prepare a replacement

sensor

ALM7

Oxygen concentration

high/low limit error

The oxygen concentration became larger than

the high/low limits.

Change the high/low limits

as needed

ALM8

Over range error

The oxygen concentration became greater than

100%FS, or became a negative value.

Change the range

Re-calibrate

7.2.2 Inspection in the Event of an Error (ERR)

If an error is detected during operation, the ERR/ALM lamp turns on and an error (ERR) code is

displayed on the sub-display. If an error is detected, take action according to Table 7.2.

Table 7.2

Error (ERR) Conditions and Actions

Display

Factor

Diagnosis Condition

Action

Err1

Sensor failure

Sensor burnout. The sensor

electromotive force became

less than -50 mV.

Replace the sensor

Err2

Heater temperature error

The heater temperature

became greater by ±30°C

relative to the standard

temperature.

Replace the heater

Check the temperature adjustment circuit

(Contact our service department)

Err3

Temperature sensor burnout The temperature sensor has

burned out.

Replace the heater

(Contact our service department)

Err4

OX400 temperature error

The temperature inside the

OX400 became greater

than 70°C.

Check to make sure the ventilation holes,

cooling fan outlet, and line are not blocked,

lines are not blocked, and so on.

Turn the power off and then on.

Err5

CPU failure

A CPU failure occurred.

Contact our service department.

Err6

Fan stop

The cooling fan stopped.

Contact our service department.