Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 16

Advertising
background image

IM 12D08B02-01E

Specifications 2-3

NEPSI Certification (SC202S-K)

NEPSI Intrinsically Safe Type

Cert No. GYJ081157X

• Applicable Standard:

GB3836.1-2000, GB3836.4-2000

• Type of Protection and Marking Code:

Ex ia IIC T4/T6

• Ambient Temperature :

T6; –10 to 40°C, T4; –10 to 55°C

Note 1 Entity Parameters

• Intrinsically safe input parameters

(terminal + and -):

Maximum Input Voltage (Ui) = 31.5 V

Maximum Input Current (Ii) = 100 mA

Maximum Input Power (Pi) = 1.2 W

Maximum Internal Capacitance (Ci) = 22 nF

Maximum Internal Inductance (Li) = 35 μH

• Intrinsically safe output parameters and

maximum external parameters

(terminal 11 and 16):

Uo=14.4 V, Io=13 mA, Po=185 mW, Co=59

nF, Lo=200 mH

Note 2 Installation

• Electrostatic charges on the display window

shall be avoided.

• The external earth connection facility shall be

connected reliably.

• The instrument modification or parts

replacement by other than authorized

representative of Yokogawa Electric

Corporation and will void NEPSI Intrinsically

safe certification.

• The user shall not change the configuration

in order to maintain/ensure the explosion

protection performance of the equipment. Any

change may impair safety.

• For installation, use and maintenance of

the product, the end user shall observe the

instruction manual and the following standards:

GB50257-1996 "Code for construction and

acceptance of electric device for explosion

atmospheres and fire hazard electrical

equipment installation engineering''.

Item

Description

Code

IECEx

Type of protection "n"

Applicable standard: IEC 60079-15:2001,

IEC 60079-0:2004

Certificate: IECEx KEM

06.0053X

Ex nA [nL] IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=31.5 V, Ci=22 nF, Li=35 μH

IECEx Intrinsically safe

Applicable standard: IEC 60079-0, IEC60079-11,

IEC60079-26

Certificate: IECEx KEM

06.0053X

Zone 0 Ex ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=31.5 V, Ii=100 mA, Pi=1.2 W, Ci=22 nF, Li=35 μH

-N

-A

T12E.EPS

IECEx

Scheme

CSA Non-incendive safe Approval or
t

ype of protection "n"

Applicable standard: C22.2, No.0-M1991,

C22.2, No.04-M2004, C22.2, No.157-M1992,

C22.2, No.213-M1987, C22.2, No.61010-1

Class I, Division 2, Groups ABCD

Ex nA [nL] IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui(Vmax)=31.5 V, Ci=22 nF, Li=35 μH

-N

CSA Intrinsically safe Approval

Applicable standard: C22.2, No. 0-M1991,

C22.2, No. 04-M2004, C22.2, No. 157-M1992,

C22.2, No. 61010-1

Ex ia Class I,

Division

1, Groups ABCD

Ex ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui(Vmax)=31.5 V, Ii(Imax)=100 mA,

Pi(Pmax)=1.2 W, Ci=22 nF, Li=35 μH

-A

Canadian

Standards

Association

(CSA)

Item

Description

Code

CENELEC ATEX (KEMA)

Type of protection "n"

Applicable standard: EN60079-0:2006,

EN60079-15:2003

Certificate: KEMA 07ATEX0051

EEx nA [nL] IIC, Group: II, Category: 3G

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=32 V, Ci=220 pF, Li=0 μH

CENELEC ATEX (KEMA) Intrinsically safe Approval

Applicable standard: EN60079-0, EN50020

EN60079-26

Certificate: KEMA 07ATEX0050 X

Ex ia IIC, Group: II, Category: 1G

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=220 pF, Li=0 μH

CENELEC

ATEX

ATEX.EPS

Entity

CENELEC

ATEX

CENELEC

ATEX

FISCO

-F

-P

or

-D

-B

or

CENELEC ATEX (KEMA) Intrinsically safe Approval

Applicable standard: EN60079-0, EN50020

EN60079-26,

EN60079-27

Certificate: KEMA 07ATEX0050 X

Ex ia IIC, Group: II, Category: 1G

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=220 pF, Li=0 μH

mA

Item

Description

Code

IEC.EPS

Entity

FISCO

-F

-P

or

-D

-B

or

IECEx

Type of protection "n"

Applicable standard: IEC 60079-15:2001,

IEC

60079-0:2004

Certificate: IECEx KEM 07.0027X

Ex nA [nL] IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=32 V, Ci=220 pF, Li=0 μH

IECEx Intrinsically safe

Applicable standard: IEC 60079-0, IEC60079-11,

IEC60079-26

Certificate: IECEx KEM 07.0027X

Zone 0 Ex ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=220 pF, Li=0 μH
IECEx Intrinsically safe

Applicable standard: IEC 60079-0, IEC60079-11,

IEC60079-26, IEC60079-27

Certificate: IECEx KEM 07.0027X

Zone 0 Ex ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=220 pF, Li=0 μH

IECEx

Scheme

IECEx

Scheme

IECEx

Scheme

mA

Item

Description

Code

CSA Non-incendive safe Approval or
t

ype of protection "n"

Applicable standard: C22.2, No.0-M1991,

C22.2, No.04-M2004, C22.2, No.157-M1992,

C22.2, No.213-M1987, C22.2, No.61010-1

Class I, Division 2, Groups ABCD

Ex nA [nL] IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

CSA Intrinsically safe Approval

Applicable standard: C22.2, No. 0-M1991,

C22.2, No. 04-M2004, C22.2, No. 157-M1992,

C22.2, No. 61010-1

Ex ia Class I,

Division

1, Groups ABCD

Ex ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

CSA.EPS

Canadian

Standards

Association

(CSA)

Ui(Vmax)=24 V, Ii(Imax)=250 mA,

Pi(Pmax)=1.2 W, Ci=220 pF, Li=0 μH

Ui(Vmax)=17.5 V, Ii(Imax)=380 mA,

Pi(Pmax)=5.32 W, Ci=220 pF, Li=0 μH

Entity

FISCO

Ui(Vmax)=32 V, Ci=220 pF, Li=0 μH

Ui(Vmax)=32 V, Ci=220 pF, Li=0 μH

Entity:

FNICO:

-F

-P

or

-D

-B

or

Advertising