IM 12D08B02-01E

2-6 Specifications

2-4. Control Drawing SC202S mA HART® Specification (IECEx)

SC 202S (C onductivity/R esistivity-transmitter)

Safe area

U o = 31.5 V o lt D C

Io = 100 mA

Po = 1.2 W att

Ex ia or ib C ertified R epeater
Pow er Supply
(HA R T C ompatible)

O utput

Supply

SE N SO R (S)
term inals 11-16

+

_

G

SEN SO R (S)
term inals 11-16

Zo ne 0 o r 1

H azardous area

Safe area

+

_

G

Functional
earth

earth

Intrinsically safe design
IEC Ex standard Ex ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

C ertificate nr. IEC Ex K EM 06.0053X
SC 202S

+

_

Load

R esistance

Ex ia or ib
C ertified safety barrier or pow er
w ith R int= 300

:

Io = 100 mA

U o = 31.5 V o lt D C

Zo ne 0 o r 1

H azardous area

earth

+

_

24 volts D C N ominal

Supply V oltage.

Intrinsically safe design

IEC Ex standard Ex ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp.< 40°C

C ertificate nr. IEC Ex K EM 06.0053X

(HA R T compatible)

(C onductivity/R esistivity-transmitter)

Functional

Functional

Sensor(s) are of a passive type to be regarded as ‘simple apparatus’.

Electrical data of the SC202S.

- Supply and output circuit (terminals + and -):

Maximum input voltage U

i

= 31.5 V. Maximum input current I

i

= 100 mA.

Maximum input power P

i

= 1.2 W.

Effective internal capacitance

C

i

= 22 nF.

Effective internal inductance

L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 16):

Maximum output voltage U

o

= 14.4 V. Maximum output current I

o

= 13 mA.

Maximum allowed external capacitance

Co = 59 nF (for SC202S-A),

Co = 2.9 PF (for SC202S-N).

Maximum allowed external inductance

Lo = 200 mH (for SC202S-A),

Lo = 450 mH (for SC202S-N).

Barriers and power supply specification must not exceed the maximum values

as shown in the diagram above. These safety descriptions cover most of

the commonly used industry standard barriers, isolators and power supplies.

The Hand Held Communicator must be of a IECEx certified intrinsically safe type

in case it is used on the intrinsically safe circuit in the hazardous area or of a

IECEx certified non-incendive type in case it is used in the non-incendive circuit

in the hazardous area.