Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 63

Advertising
background image

IM 12D08B02-01E

Parameter setting 5-17

Code Display

Function

Function

detail

X

Y

Z

Default

values

User interface
50

*RET.

Auto return

Auto return to measuring mode Off

0

Auto return to measuring mode On

1

1

On

51

Not

used

52 *PASS Passcode

Maintenance

passcode

Off

0 0.0.0

Off

Note # = 0 - 9, where Maintenance passcode On

#

Commissioning passcode Off

0

Off

1=111, 2=333, 3=777 Commissioning passcode On

#

4=888, 5=123, 6=957 Service passcode Off

0

Off

7=331, 8=546, 9=847 Service passcode On

#

53

*Err.01

Error setting

Polarization too high

Soft/Hard

0/1

1

Hard

*Err.05

Shorted

measurement

Soft/Hard

0/1

1

Hard

*Err.06

Open

measurement Soft/Hard

0/1

1

Hard

*Err.07

Temperature sensor open Soft/Hard

0/1

1

Hard

*Err.08

Temp. sensor shorted

Soft/Hard

0/1

1

Hard

*Err.13

USP limit exceeded

Soft/Hard

0/1

0

Soft

54

*E5.LIM

E5 limit setting

Maximum conductivity value

250 mS

(Minimum resistivity value)

0.004 kΩ

*E6.LIM

E6 limit setting

Minimum conductivity value

1.000 μS

(Maximum resistivity value)

1.000 MΩ

55

*%

Display mA in w/w% mA-range displayed in w/w% off

0

0

Off

mA-range displayed in w/w% on

1

*0%

Set 0% output value in w/w%

*100%

Set 100% output value in w/w%

56

*DISP.

Display resolution

Auto ranging display

0

0

Auto

Display fixed to X.XXX μS/cm or MΩ·cm 1

Display fixed to XX.XX μS/cm or MΩ·cm 2

(2)

Display fixed to XXX.X μS/cm or MΩ·cm 3

Display fixed to X.XXX mS/cm or kΩ·cm 4

Display fixed to XX.XX mS/cm or kΩ·cm 5

Display fixed to XXX.X mS/cm or kΩ·cm 6

Display fixed to XXXX mS/cm or kΩ·cm 7

57

*USP.

USP setting

Disable the E13 (USP limit exceeded)

0

0

Off

Enable the E13 (USP limit exceeded)

1

58-59

Not

used