Operation 4-3

4-3. Setting passcodes

4-3-1. Passcode protection

In Service Code 52, EXA users can set passcode protection for each one of the three operating levels,

or for any one or two of the three levels. This procedure should be completed after the initial commis-

sioning (setup) of the instrument. The passcodes should then be recorded safely for future reference.

When passcodes have been set, the following additional steps are introduced to the configuration and

programming operations:

Maintenance

Press MODE key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Maintenance Mode

Commissioning

Press * key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Commissioning Mode.

Service

From the commissioning menu, select *SERV by pressing YES key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Service Mode.

NOTE

:

See Service Code 52 for the setting of passcodes.

4-4. Display examples

The following pages show the sequence of button presses and screens displayed when working in some

standard configurations. More or less options will be made available by the configuration of some service

codes, or by choices made in the commissioning menu.

The following deviations are possible:

Item marked is omitted when switched off in commissioning mode.

Temperature compensation will be displayed dependent on chosen compensation method: NaCl,

TC or matrix.

DISP.2 only appears if a 2nd (different) temperature compensation (*TEMP.2) is set.

W/W % only appears if switched on in service code 55. In display 2 w/w % never appears.