Calibration, 1 when is calibration necessary, Calibration -1 – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 66: 1 when is calibration necessary? -1

Advertising
background image

IM 12D08B02-01E

Calibration 6-1

6. CALIBRATION

6-1 When is calibration necessary?

Calibration of conductivity/resistivity instruments is normally not required, since Yokogawa delivers a

wide range of sensors, which are factory calibrated traceable to NIST standards. The cell constant

values are normally indicated on the top of the sensor or on the integral cable. These values can be

entered directly in service code 03 (section 5-3-1). If the cell has been subjected to abrasion (erosion or

coating) calibration may be necessary. In the next section two examples are given. Alternatively calibra-

tion may be carried out with a simulator to check the electronics only.

NOTE

:

During calibration the temperature compensation is still active. This means that the readings are

referred to the reference temperature as chosen in service code 20 (section 5-4, default 25 °C).

Calibration is normally carried out by measuring a solution with a known conductivity value at

a known temperature. The measured value is adjusted in the calibration mode. On the next

pages the handling sequence for this action is visualized. Calibration solutions can be made up

in a laboratory. An amount of salt is dissolved in water to give a precise concentration with the

temperature stabilized to the adjusted reference temperature of the instrument (default 25 °C). The

conductivity of the solution is taken from literature tables or the table on this page.

Alternatively the instrument may be calibrated in an unspecified solution against a standard instrument.

Care should be taken to make a measurement at the reference temperature since differences in the type

of temperature compensation of the instrument may cause an error.

NOTE

:

The standard instrument used as a reference must be accurate and based on an identical

temperature compensation algorithm. Therefore the Model SC72 Personal Conductivity Meter of

Yokogawa is recommended.

Typical calibration solutions.

The table shows some typical conductivity values for sodium-chloride (NaCl) solutions which can be

made up in a laboratory.

Table 6-1. NaCl values at 25°C (IEC 60746-3)

Weight %

mg/kg

Conductivity

0.001

10

21.4 μS/cm

0.003

30

64.0 μS/cm

0.005

50

106 μS/cm

0.01

100

210 μS/cm

0.03

300

617 μS/cm

0.05

500

1.03 mS/cm

0.1

1000

1.99 mS/cm

0.3

3000

5.69 mS/cm

0.5

5000

9.48 mS/cm

1

10000

17.6 mS/cm

3

30000

48.6 mS/cm

5

50000

81.0 mS/cm

10

100000

140 mS/cm

NOTE

:

For resistivity measurement the standard resistivity

units of the calibration solution can be calculated

as follows:

R = 1000/G kΩ·cm (if G = μS/cm)

Example:

0.001 weight %

R = 1000/21.4 = 46.7 kΩ·cm

Advertising