Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 59

Advertising
background image

IM 12D08B02-01E

Parameter setting 5-13

Code Display

Function

Function

detail

X

Y

Z

Default

values

Temperature compensation functions
20

*T.R.°C

Set reference temp.

Use >, ^, ENT keys to set value

25 °C

21

*T.C.1

Set temp. coef. 1

Adjust compensation factor

2.1 %

if set to TC in section 5-2-5.

per °C

Set value with >, ^, ENT keys

*T.C.2

Set temp. coef. 2

Adjust compensation factor

2.1 %

if set to TC in section 5-2-5.

per °C

Set value with >, ^, ENT keys

22

*MATRX

Select matrix

Choose matrix if set to matrix comp.

in section 5-2-5, using >, ^, ENT keys

HCl (cation) pure water (0-80 °C)

1

1

HCI

Ammonia pure water (0-80 °C)

2

Morpholine pure water (0-80 °C)

3

HCl (0-5 %, 0-60 °C)

4

NaOH (0-5 %, 0-100 °C)

5

User programmable matrix

9

23

*T1 °C (°F) Set temp. range

Enter 1st (lowest) matrix temp. value

*T2..

Enter 2nd matrix temp. value

*T3..

Enter 3rd matrix temp. value

*T4..

Enter 4th matrix temp. value

*T5..

Enter 5th (highest) matrix temp. value

24

*L1xT1

Enter conductivity

Value for T1

*L1xT2

values for lowest

Value for T2

....

concentration

*L1xT5

Value for T5

25

*L2xT1

Concentration 2

Similar to code 24

26

*L3xT1

Concentration 3

Similar to code 24

27

*L4xT1

Concentration 4

Similar to code 24

28

*L5xT1

Concentration 5

Similar to code 24

29

Not

used