3. calibration with hold active, 3. calibration with hold active -3 – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 68

Advertising
background image

IM 12D08B02-01E

Calibration 6-3

6-3. Calibration with HOLD active

HOLD

ENT

ENT

>

Select the flashing digit with the > key.

Increase its value by pressing the key

>

After the indication is stable,

set the value

using the >, , ENT key.

Put the sensor in standard

solution. Press YES.

When the correct value is displayed,

press ENT to enter the change.

After briefing displaying WAIT,

the CAL.END message appears.

The calibration is now complete. Put the

sensor back in the process and press YES.

HOLD will be displayed. Press NO to turn off

HOLD and return to the measuring mode.

NO

YES

NO

YES

NO

YES

HOLD

MODE

NO

YES

YES

HOLD

ENT

NO

MODE

YES

MODE

MEASURE
CAL
DISPLAY

HOLD

Press the MODE key.

The legend CALIB

appears, and the YES/NO

key prompt flags flash.

HOLD

HOLD

ENT

HOLD

HOLD