Connections differential 4-electrode, C-13 d-14 – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 41

Advertising
background image

IM 12D08B02-01E

3-8 Installation and wiring

14 Overall Screen

11

12

12

13

14

14

16

15

13

14

14

16

15

17

11

17

11 Red

12 Blue

15 Core 16 Screen
White Co-axial cable

13 Core 17 Screen
Brown Co-axial Cable

WF10 Cable

TRANSMITTER / CONVERTER

A

C

Overall

shield

B

D

Screen

E

Red

White

Blue

Brown

Thermistor (Temperature sensor)

Ground (Shield)

Secondary Coil

Primary Coil

11
12
17
13
15
16
14

Fig. 3-12. Connection of WF10 extension cable and BA10/BP10 junction box

NOTE

:

See page 3-10 for termination for WF10 cable in combination with EXA SC

A-15
B-16

E-11
F-12
S-3 or 63

temp

>Connections differential 4-electrode

C-13
D-14

Advertising