Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 24

Advertising
background image

IM 12D08B02-01E

Specifications 2-11

2-9. Control Drawing of SC202S FF/PB Specification (IECEx)

Safe area

Hazardous area

Zone 0 or 1

Safe area

Apparatus

I.S.

interface

I.S.

certified

Terminator

Ui = 24 V

or Ui = 17,5 V

Ii = 250 mA Ii = 380 mA

Pi = 1,2 W Pi = 5,32 W

Ex ia IIC

T4 for ambient temp. d 55 qC

SC202S-F

or SC202S-P

I.S.

certified

Terminator

-

+

-

+

Transmitter

Transmitter

Sensor

Connections

-

+

-

+

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus'.

x

Electrical data of the SC202S-F & SC202S-P:
-

Supply

and

output

circuit:

Maximum input voltage Ui= 24 V

Maximum input current Ii= 250 mA

Maximum input power Pi= 1.2 W

Effective internal capacitance Ci= 220 pF;

Effective internal inductance Li= 0 μH.

or

FISCO

field

device

Maximum input voltage Ui= 17.5 V

Maximum input current Ii= 380 mA

Maximum input power Pi= 5.32 W

Effective internal capacitance Ci= 220 pF;

Effective internal inductance Li= 0 μH.

-

Sensor input circuit:

Maximum output voltage Uo= 14.4 V; Maximum output current Io= 13 mA

Maximum allowed external capacitance Co= 59 nF

Maximum allowed external inductance Lo= 200 mH

x

Any I.S. interface may be used that meets the following requirements:

Uo d 24 V

Io d 250 mA

Po d 1.2 W

Ca t 220 pF + Ccable; La t 0 μH + Lcable

or

FISCO power supply

Uo d 17.5 V

Io d 380 mA

Po d 5.32 W

Ca t 220 pF + Ccable; La t 0 μH + Lcable

x

Electrical data of the SC202S-B & SC202S-D (Type of protection “n”)
-

Supply

and

output

circuit:

Maximum input voltage Ui = 32 V

Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 μH.

-

Sensor input circuit:

Maximum output voltage Uo= 14.4 V; Maximum output current Io= 13 mA

Maximum allowed external capacitance Co = 2.9 μF

Maximum allowed external inductance Lo = 450 mH

Advertising