IM 12D08B02-01E

Specifications 2-13

2-11. Control Drawing of SC202S FF/PB Specification (FM Intrinsically safe Entity)

Unclassified Location

Classified Location

Division 1

FM Approved

barrier

Voc (Vt) d 24 V

Ioc (It) d 250 mA

Poc (Pt) d 1.2 W

Ca t 220pF+ Ccable

La t 0 H + Lcable

Sensor

Connections

Max. cablelength: 60 mtr.

Cable dia. : 3…12 mm.

FM Class I, DIV. 1, Group ABCD

T4 for ambient temp. d 55 qC

-

+

I.S.

certified

Terminator

I.S.

certified

Terminator

SC202S-F

or SC202S-P

Transmitter

Transmitter

-

+

Sensor

Connections

-

+

-

+

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which

neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or

energy over 20 PJ, or are FM Approvals entity approved and meet connection

requirements.

x

Electrical data of the SC202S-F & SC202S-P:
-

Supply

circuit:

Maximum input voltage Vmax= 24 V

Maximum input current Imax= 250 mA

Maximum input power Pi= 1.2 W

Effective internal capacitance Ci = 220 pF;

Effective internal inductance Li = 0 PH.

-

Sensor input circuit:

Maximum output voltage Vt= 14.4 V; Maximum output current It= 10 mA

Maximum allowed external capacitance Ca= 59.36 nF

Maximum allowed external inductance La= 340 mH

x

Any FM Approved barrier may be used that meets the following requirements:

Voc or Vt d 24 V

Ioc or It d 250 mA

Poc or Pt d 1.2 W

Ca t 220 pF + Ccable; La t 0 PH + Lcable

When installing this equipment, follow the manufacturer’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of

Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National

Electrical Code (ANSI/NFPA 70).

Associated apparatus connected to the barrier must not use or generate more than

250 Vrms or Vdc.

x

Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0

Ohm.

x

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

WARNING

- Substitution of components may impair Intrinsic Safety

- To prevent ignition of flammable or combustible atmospheres, disconnect power

before servicing or read, understand and adhere to the manufacturer’s live

maintenance procedures.