Installation and wiring 3-7

3-6. Other sensor systems

To connect other sensor systems, follow the general pattern of the terminal connections as listed below:

11 and 12

: Always used for temperature compensation resistor input.

13 and 14

: Normally used for the outer electrode

15 and 16

: Used for inner electrode

In case a 4-electrode measuring system will be used, 14 and 16 should be used for the current elec-

trodes.

Please ensure that shielded cabling will be used.

In figure 3-10 this is shown in a schematic way.

Figure 3-10. Connection diagram for other sensors

Figure 3-11. Terminal identification label

3-6-1. Sensor cable connections using junction box (BA10) and extension cable (WF10)

Where a convenient installation is not possible using the standard cables between sensors and transmit-

ter, a junction box and extension cable may be used. The Yokogawa BA10 junction box and the WF10

extension cable should be used. These items are manufactured to a very high standard and are neces-

sary to ensure that the specifications of the system are not compromised. The total cable length should

not exceed 60 metres (e.g. 5 m fixed cable and 55 m extension cable).

Note

: 17 of both WF10 and BA10 do not need to be used.

11 12 13 14 15 16

t

11 12 13 14 15 16

t

2-electrode configuration

4-electrode configuration