Installation and wiring 3-1

3. INSTALLATION AND WIRING

3-1. Installation and dimensions

3-1-1. Installation site

The EXA transmitter is weatherproof and can be installed inside or outside. It should, however, be

installed as close as possible to the sensor to avoid long cable runs between sensor and transmitter. In

any case, the cable length should not exceed 60 meters (200 feet). Select an installation site where:

• Mechanical vibrations and shocks are negligible

• No relay/power switches are in the direct environment

• Access is possible to the cable glands (see figure 3-1)

• The transmitter is not mounted in direct sunlight or severe weather conditions

When the instrument with Suffix Code "-B,-N,-D" is used, take measures so that the display window is

not exposed to direct sunlight

• Maintenance procedures are possible (avoiding corrosive environments)

The ambient temperature and humidity of the installation environment must be within the limits of the

instrument specifications. (See chapter 2).

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the EXA transmitter has universal mounting capabilities:

• Panel mounting using optinal bracket, refer to Fig. 3-2a.

• Panel mounting using two (2) self-tapping screws, refer to Fig. 3-2b.

• Surface mounting on a plate (using bolts from the back)

• Wall mounting on a bracket (for example, on a solid wall)

• Pipe mounting using a bracket on a horizontal or vertical pipe (nominal pipe diameter JIS 50A)

Fig. 3-2a. Panel mounting diagram

Fig. 3-1. Housing dimensions and layout of glands

162

202

155

180

30

77

50

42

68

115

130

56±0.2

Hood (Option)
Option code : / H

Grounding terminal

(M4 screw)

Transmission signal cable inlet

Cable gland : Pg13.5

Sensor cable inlet

cable gland (Pg13.5)

M6 screw (32 )

1.eps

9

34

(7.95)

(6.4)

(6.1)

(7)

(2)

(5.1)

(1.2)

(3)

(4.5)

(1.65)

(2.7)

(0.35)

(1.26) depth

(1.3)

(2.20)

1 to 10

205

185

3.9

80

38

60

30

Panel thickness

Panel mounting

bracket

173

+1.1

0

156

+1.1

0

121

138

PANEL CUTOUT

5.eps

(0.04 to 0.39)

(7.28)

(8.07)

(1.5) (3.15) (0.15)

(6.14)

(4.76)

(6.81) (5.43)

(1.2)

(2.36)

49

Adapter

G1/2 screw (/AFTG)

1/2 NPT screw (/ANSI)

Approx.

55

2.eps

(2.2)

(1.93)

Unit: mm (inch)

Advertising