4. temperature compensation functions, 4. temperature compensation functions -12 – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 58

Advertising
background image

IM 12D08B02-01E

5-12 Parameter setting

5-4. Temperature compensation functions

Code 20

*T.R.°C

Choose a temperature to which the measured conductivity (or resistivity) value

must be compensated. Normally 25°C is used, therefore this temperature is

chosen as default value. Limitations for this setting are: 0 to 100 °C.

If *T.UNIT in code 11 is set to °F, default value is 77°F and the limitations are

32 - 212°F.

Code 21

*T.C.1, 2

In addition to the procedure described in section 5-2-4 it is possible to adjust

the compensation factor directly. If the compensation factor of the sample liquid

is known from laboratory experiments or has been previously determined, it

can be introduced here.

Adjust the value between 0.00 to 3.50 % per °C. In combination with reference

temperature setting in code 20 a linear compensation function is obtained,

suitable for all kinds of chemical solutions.

Code 22

*MATRX

The EXA is equipped with a matrix type algorithm for accurate temperature

compensation in various applications. Select the range as close as possible

to the actual temperature/concentration range. The EXA will compensate by

interpolation and extrapolation. Consequently, there is no need for a 100%

coverage.

If 9 is selected the temperature compensation range for the adjustable matrix

must be configured in code 23. Next the specific conductivity values at the

different temperatures must be entered in codes 24 to 28.

Code 23

*T1, T2, T3, Set the matrix compensation range. It is not necessary to enter equal

T4 & T5 °C temperature steps, but the values should increase from T1 to T5, otherwise

the entrance will be refused. Example: 0, 10, 30, 60 and 100 °C are valid

values for the T1....T5. The minimum span for the range (T1 - T5) is 25 °C.

Code 24-28 *L1xT1 - In these access codes the specific conductivity values can be entered for

L5xT5

5 different concentrations of the process liquid; each one in one specific access

code (24 to 28). The table below shows a matrix entering example for 1 - 15%

NaOH solution for a temperature range from 0 - 100 °C.

NOTES

:

1. In chapter 11 a table is included to record your programmed values. It will make programming easy

for duplicate systems or in case of data loss.

2. Each matrix column has to increase in conductivity value.

3. Error code E4 occurs when two standard solutions have identical conductivity values at the same

temperature within the temperature range.

Table 5-2. Example of user adjustable matrix

Matrix

Example

Example

Example

Example

Example

Code 23

Temperature

T1...T5

0 °C

25 °C

50 °C

75 °C

100 °C

Code 24

Solution 1 (1%)

L1

31 mS/cm

53 mS/cm

76 mS/cm

98 mS/cm

119 mS/cm

Code 25

Solution 2 (3%)

L2

86 mS/cm

145 mS/cm

207 mS/cm

264 mS/cm

318 mS/cm

Code 26

Solution 3 (6%)

L3

146 mS/cm

256 mS/cm

368 mS/cm

473 mS/cm

575 mS/cm

Code 27

Solution 4 (10%)

L4

195 mS/cm

359 mS/cm

528 mS/cm

692 mS/cm

847 mS/cm

Code 28

Solution 5 (15%)

L5

215 mS/cm

412 mS/cm

647 mS/cm

897 mS/cm

1134 mS/cm