3-4 Installation and wiring

Figure 3-6. System configuration

3-3. Wiring of sensors

3-3-1. General precautions

Generally, transmission of signals from SC sensors is at a low voltage and current level. Thus a lot of

care must be taken to avoid interference. Before connecting sensor cables to the transmitter make sure

that following conditions are met:

– the sensor cables are not mounted in tracks together with high voltage and or power switching cables

– only standard sensor cables or extension cable are used

– the transmitter is mounted within the distance of the sensor cables (max. 10 m) + up to 50 m WF10

extension cable.

– the setup is kept flexible for easy insertion and retraction of the sensors in the fitting.

3-3-2. Additional precautions for installations in hazardous areas - Intrinsic safe

Make sure that the total of capacitance and inductances connected to the input terminals of the EXA

SC202S do not exceed the limits given in the certificate.

This sets a limit to the cable and extensions used.

– The intrinsic safe version of the EXA 202 instrument can be mounted in Zone 1.

– The sensors can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in the

system certificate is used.

– Ensure that the total of capacitance and inductances connected to the terminals of the EXA SC202 do

not exceed the limits given in the certificate of the safety barrier or distributor.

– The cable used should preferably have a BLUE colour or marking on the outside.

– Installation for (sensors in Zone 0 or 1):

Generally, the distributor with input/output isolation has no external earth connection. If there is an

earth connection on the distributor and the external connection of the transmitter is connected to

“protective” earth, the shield of the 2-wire cable may NOT be connected to “protective” earth at the

distributor too.

INPUT

OUTPUT/SUPPLY

Safety Barrier

SC202S only

SENSORS

2,5 or 10 m

DISTRIBUTOR

RECORDER

COMPUTER

HAND HELD

COMMUNICATOR

CURRENT OUTPUT

NO

MODE

YES

ENT

>

FAIL

HOLD

YES NO

ENT

MEASURE

MAN.CAL
DISPLAY

HOLD

YOKOGAWA

MODE

TEMP

AUT.CAL

1 2

1 8 0

100

0

>

Advertising