11-7. Error codes

Code Error description

Possible cause

Suggested remedy

E1

Polarization detected on cell

Sensor surface fouled

Clean sensor

Conductivity too high

Replace sensor

E2

Temperature coefficient out of limits

Incorrect field calibration of TC

Re-adjust

(0-3.5%/ºC)

Set

calculated

TC

E3

Calibration out of limits

Calibrated value differs more than

Check for correct sensor

+/- 20 % of nominal value programmed Check for correct unit (μS/cm,

in code 03.

mS/cm, kΩ·cm or MΩ·cm)

Repeat

calibration

E4

Matrix compensation error

Wrong data entered in 5x5 matrix

Re-program

E5

Conductivity too high or resistivity too low

Incorrect wiring

Check wiring (3-6)

(Limits set in service code 54)

Internal leakage of sensor

Replace sensor

Defective cable

Replace cable

E6

Conductivity too low or resistivity too high

Dry sensor

Immerse sensor

(Limits set in service code 54)

Incorrect wiring

Check wiring (3-6)

Defective cable

Replace cable

E7

Temperature sensor open

Process temperature too high or too low Check process

(Pt1000Ω : T > 250°C or 500°F)

Wrong sensor programmed

Check model code sensor

(Pt100Ω/Ni100Ω : T > 200°C or 400°F)

Incorrect wiring

Check connections and cable

(8k55 : T < -10°C or 10°F)

(PB36 : T < -20°C or 0°F)

E8

Temperature sensor shorted

Process temperature too high or too low Check process

(Pt1000Ω/Pt100Ω/Ni100Ω : T < -20°C or 0°F) Wrong sensor programmed

Check model code sensor

(8k55/PB36 : T > 120°C or 250°F)

Incorrect wiring

Check connections and cable

E9

Air set impossible

Too high zero due to cable capacitance

Replace cable

E10 EEPROM write failure

Fault in electronics

Try again, if unsuccessful con-
tact Yokogawa

E13 USP limit exceeded

Poor water quality

Check ion exchangers

E15 Cable resistance influence to temperature

Cable resistance too high

Check cable

exceeds +/- 15°C

Corroded contacts

Clean and reterminate

Wrong sensor programmed

Reprogram

E17 Output span too small

Incorrect configuration by user

Reprogram

E18 Table values make no sense

Wrong data programmed

Reprogram

E19 Programmed values outside acceptable limits Incorrect configuration by user

Reprogram

E20 All programmed data lost

Fault in electronics

Contact Yokogawa

Very severe interference

E21 Checksum error

Software problem

Contact Yokogawa