Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 23

Advertising
background image

IM 12D08B02-01E

2-10 Specifications

Sensor is a thermocouples, RTD’s, passive resistive switch devices, or is CSA entity

approved and meet connection requirements.

Electrical data of the SC202S.

- Supply and output circuit (terminals + and -)

Maximum input voltage Vmax = 31.5 V. Maximum input current Imax = 100 mA.

Maximum input power Pmax = 1.2 W.

Effective internal capacitance Ci = 22 nF. Effective internal inductance Li = 35 PH.

- Sensor input circuit (terminals 11 through 16):

Maximum output voltage Voc = 14.4 V. Maximum output current Isc = 13 mA.

Maximum allowed external capacitance Ca = 59 nF.

Maximum allowed external inductance La = 200 mH.

Barriers and power supply should be CSA certified. The specifications must not exceed

the maximum values as shown in the diagram above. Installation should be in

accordance with Canadian Electrical Code, Part I.

Maximum safe area voltage should not exceed 250 VRMS.

For Class I, Div. 2, Group ABCD the CSA certified barrier is not required, and the Sensor

input circuit (terminals 11 through 16) is non-incendive having the parameters :

Maximum output voltage Voc = 14.4 V. Maximum output current Isc = 13 mA.

Maximum allowed external capacitance Ca = 2.9 PF.

Maximum allowed external inductance La = 450 mH.

The Hand Held Communicator must be of a CSA certified intrinsically safe type in case

it is used on the intrinsically safe circuit in the hazardous area, or of a CSA certified

non-incendive type in case it is used on the non-incendive circuit in the hazardous area.

S a fe a re a

V m ax = 3 1 .5 V o ltD C
Im ax = 1 0 0 m A

P m ax = 1 .2 W att

C S A c e rtifie d
P o w e r S up p ly
(H A R T c o m p atib le )

O utp ut

S up p ly

S en so r
term in als 1 1 -1 6

+

_

G

S en so r
term in als 1 1 -1 6

H a z ard o us a re a

S a fe are a

+

_

G

e arth

e a rth

Intrins ica lly safe d esign

transm itte r

S C 2 0 2 S

+

_

Lo a d

R e sis ta nc e

Im ax = 1 0 0 m A

V m ax = 3 1 .5 V o ltD C

H a z ard o us a re a

ea rth

+

_

2 4 vo lts D C N o m ina l

S up p ly V o lta ge .

Intrins ica lly safe d esign

C S A E x ia C lass1 , D iv.1 , G ro u p A B C D , T 4 fo r am bient tem p . < 5 5 °C
T 6 fo r am bient tem p . < 4 0 °C

S C 2 0 2 S transm itte r

C S A certified

sa fe ty b a rrie r o r p o w e r sup p ly
w ith R int= 3 0 0

:

S u itable valu es are:

(H A R T c o m p a tib le )

C S A E x ia C lass1 , D iv.1 , G ro u p A B C D , T 4 fo r am bient tem p . < 5 5 °C
T 6 fo r am bient tem p. < 4 0 °C

Fo r electrical d ata:
see text belo w .

Fo r electrical d ata:
see text belo w .

S uitab le va lue s a re :

Functio na l

Func tio na l

Func tio na l

2-8. Control Drawing of SC202S mA HART® Specification (CSA)

Advertising