(9) After the test at 20 mA, press the [ENT] key twice.

Table 1

Current output (mA)

Current output (mA DC)

4.0 4

±0.02

12.0 12

±0.02

20.0 20

±0.02

3.3 Temperature

Indication

Check

(1) Check that the temperature sensor type indicated on the message display is “Pt1000.”

(2) Change the value of the decade resistance box 1 (temperature) as shown in Table 2 and

check the data display. The temperature value on the data display must be within the range

shown in Table 2.

(3) Press the [ENT] key until the message display shows “8k55.”

(4) Press the [ENT] key. A “*WAIT” message will flash.

Table 2

Decade box 1 resistance (:)

Data display (°C)

960.9

-10

±0.3

1289.8

75

±0.3

1721.6

190

±0.3

1904.6 (*)

240 ±0.3

(*) This item is checked under measurement mode.

3.4 Resistance (Conductivity) Indication Check

(1) Check that in the message display “*WAIT” has disappeared and instead “RES.1” appears.

(2) Set decade resistance box 2 to 10 : and check the data display. The resistance must be

within the range shown in Table 3.

(3) Press the [ENT] key. The message display will show “RES.2.”

(4) Set decade resistance box 2 to 100 : and check the data display. The resistance must

be within the range shown in Table 3.

(5) Press the [ENT] key. The message display will show “RES.3.”

(6) Set decade resistance box 2 to 1 k: and check the data display. The resistance must be

within the range shown in Table 3.

(7) Press the [ENT] key. The message display will show “RES.4.”

(8) Set decade resistance box 2 to 10 k: and check the data display. The resistance must

be within the range shown in Table 3.

(9) Press the [ENT] key. The message display will show “RES.5.”

(10) Set decade resistance box 2 to 100 k: and check the data display. The resistance must

be within the range shown in Table 3.

(11) Press the [ENT] key. The message display will show “READY.”

(12) Press the [ENT] key to restart the transmitter.
This completes all the tests.

Table 3

Message display

Decade box 2 resistance

Data display

RES.1

10 :

10.00

±0.05

:

RES.2

100 :

100.0

±0.5

:

RES.3

1 k:

1.000

±0.005

k:

RES.4

10 k:

10.00

±0.05

k:

RES.5

100 k:

100.0

±0.5

k: