5 v. maximum input current i, 100 ma. maximum input power p, 2 w. effective internal capacitance c – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 20: 22 nf. effective internal inductance l, 4 v. maximum output current i

Advertising
background image

IM 12D08B02-01E

Specifications 2-7

2-5. Control Drawing SC202S mA HART® Specifi cation (ATEX)

SC 202S (C onductivity/Resistivity-transmitter)

Safe area

U o = 31.5 V o lt D C

Io = 100 mA

Po = 1.2 W att

EEx ia or ib C ertified R epeater
Pow er Supply
(H AR T C ompatible)

O utput

Supply

SE N SO R (S)
term inals 11-16

+

_

G

SEN SO R (S)
term inals 11-16

H azardous area

Safe area

+

_

G

earth

earth

Intrinsically safe design
C EN ELEC standard EEx ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

C ertificate nr. KEM A 06ATEX 0220 X

SC 202S

+

_

Load

R esistance

EEx ia or ib
C ertified safety barrier or pow er
w ith R int=300

:

Io = 100 mA

U o = 31.5 V o lt D C

H azardous area

+

_

24 volts D C N ominal

Supply V oltage.

Intrinsically safe design
C EN ELEC standard EEx ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp.< 40°C

C ertificate nr. KEM A 06ATEX 0220 X

(HA R T compatible)

(C onductivity/R esistivity-transmitter)

Zone 0 or 1

Zone 0 or 1

Functional

Functional

earth

Functional

Sensor(s) are of a passive type to be regarded as ‘simple apparatus’.

Electrical data of the SC202S.

- Supply and output circuit (terminals + and -):

Maximum input voltage U

i

= 31.5 V. Maximum input current I

i

= 100 mA.

Maximum input power P

i

= 1.2 W.

Effective internal capacitance

C

i

= 22 nF.

Effective internal inductance

L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 16):

Maximum output voltage U

o

= 14.4 V. Maximum output current I

o

= 13 mA.

Maximum allowed external capacitance

Co = 59 nF (for SC202S-A),

Co = 2.9 uF (for SC202S-N).

Maximum allowed external inductance

Lo = 200 mH (for SC202S-A),

Lo = 450 mH (for SC202S-N).

Barriers and power supply specification must not exceed the maximum values as

shown in the diagram above. These safety descriptions cover most of the commonly

used industry standard barriers, isolators and power supplies.

The Hand Held Communicator must be of a ATEX certified intrinsically safe type in

case it is used on the intrinsically safe circuit in the hazardous area or of a ATEX

certified non-incendive type in case it is used in the non-incendive circuit in the

hazardous area.