CONFIGURATION CHECKLIST FOR SC202

Primary choices

default

alternatives

reference on page

menu

Measurement

Conductivity

Resistivity

5.8- 5.9

SC 01

Range

0-1000 μS/cm

max. 1999 mS/cm

5.3

“range”

Temperature unit

Celsius

Fahrenheit

5.10- 5.11

SC 11

Sensor

Cell constant

0.1 /cm

any value between 0.08

and 50

5.8-5.9, 6.1- 6.3

SC 03

Sensor type

2-electrode

4- electrode

5.8- 5.9

SC 02

Temperature compensator

Pt1000

Ni100, Pt100, 8k55, Pb36

5.10-5.11

SC 10

Choices
Communication

enabled

disable HART

(R)

, PH201*B

5.19

SC 60- 62

Burn out

inactive

HI or LO output on fail

5.14- 5.15

SC 32

Temperature compensation

NaCl in water

fixed T.C., matrix

5.12, 5.13, 5.5

SC 20- 28; “temp”

USP functionality

inactive

Fail if USP limits are

9.1, 9.2, 5.17

SC 57

exceeded

HOLD during maintenance

inactive

HOLD last value or fixed

value

5.17, 5.3- 5.4

“hold”, SC 50

Calibration temperature

inactive

adjustment +/- 15 °C

5.11

SC 12

ZERO calibration

inactive

adjustment +/-1 μS/cm

5.9

SC 04

Diagnostics

hard alarm on hard or soft choices

5.17

SC 53

all errors

Cell fouling alarm

active

except E13 inactive

5.9

SC 05

Password protection

inactive

password for different levels

5.17

SC 52

Output in Concentration units

inactive

linearization of output, w%

5.14 - 5.17

SC 31/35/55

on LCD