Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 30

Advertising
background image

IM 12D08B02-01E

Specifications 2-17

SC202S-B

or SC202S-D

Transmitter

-

+

Transmitter

-

+

Unclassified Location

Classified Location

Division 2

FM Approved

Power Supply

Voc d 32 VDC

Sensor

Connections

Max. cablelength: 60 mtr.

Cable dia.: 3…12 mm.

FM Class I, DIV. 2, Group ABCD

T4 for ambient temp. d 55 qC

T6 for ambient temp. d 40 qC

Sensor

Connections

-

+

-

+

FM Approved

Terminator

R = 90..100Ω

C = 0..2,2 μF

FM Approved

Terminator

R = 90..100Ω

C = 0..2,2 μF

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which

neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or

energy over 20 μJ, or are FM Approvals entity approved and meet connection

requirements.

x

Electrical data of the SC202S-B & SC202S-D:
-

Supply circuit: Vmax= 32 V; Pi= 1.2 W; Ci= 220 pF; Li= 0 PH

-

Sensor input circuit: Vt= 14.4 V; It= 10 mA; Ca= 1.71 μF; La= 600 mH

When installing this equipment, follow the manufacturers installation drawing.

Installation shall be in accordance with Article 501.4(B) of the National Electrical Code

(ANSI/NFPA 79). Nonincendive field wiring may be installed in accordance with Article

501.4(B)(3)

x

Grounding shall be in accordance with Article 250 of the National Electrical code.

x

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

WARNING

- Substitution of components may impair suitability for Division 2.

- Do not remove or replace while circuit is live unless area is know to be non-hazardous

- Explosion Hazard – Do not disconnect equipment unless area is know to be

non-hazardous

- Do not reset circuit breaker unless power has been removed from the equipment or the

area is know to be non-hazardous

2-13. Control Drawing of SC202S FF/PB Specification (FM Non-incendive Entity)

Application Doc. No.: IKE027-A10 P.9