5. sensor wiring – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 39

Advertising
background image

IM 12D08B02-01E

3-6 Installation and wiring

3-5. Sensor wiring

Refer to figure 3-9, which includes drawings that outline sensor wiring.

For the SC4AJ, SC8SG and SC210G sensors, see Appendix 2.

The EXA SC202 can be used with a wide range of commercially available sensor types if provided with

shielded cables, both from Yokogawa and other manufacturers. The sensor systems from Yokogawa fall

into two categories, the ones that use fixed cables and the ones with separate cables.

To connect sensors with fixed cables, simply match the terminal numbers in the instrument with the iden-

tification numbers on the cable ends.

The separate sensors and the WU40-LH□□ cables are also numbered, but the numbers do not always

match with the terminal numbers in the instrument. Figure 3-9 indicates how to connect the different sen-

sor types.

11 TEMPERATURE

12 TEMPERATURE

13 CELL

14 CELL

15 CELL

16 CELL

CONDUCTIVITY / RESISTIVITY TRANSMITTER

SEPARATE SENSORS WITH WU40-LH . . CABLE

11 TEMPERATURE

12 TEMPERATURE

13 OUTER ELECTRODE

14 OUTER ELECTRODE

15 INNER ELECTRODE

16 INNER ELECTRODE

SC4A... SENSORS WITH INTEGRATED CABLE

RED

11 TEMPERATURE

12 TEMPERATURE

13 OUTER ELECTRODE

14 OUTER ELECTRODE

15 INNER ELECTRODE

16 INNER ELECTRODE

YELLOW / GREEN

BROWN

BROWN

1
2

1

2

SX42-SX . . - . F SENSORS

Figure 3-9. Sensor wiring diagrams