Appendix 3-7

3/3

QIS 12D08B02-21E

Table 3

Message display

Decade box 2 resistance

Data display

RES.1

10 :

10.00

±0.05

:

RES.2

100 :

100.0

±0.5

:

RES.3

1 k:

1.000

±0.005

k:

RES.4

10 k:

10.00

±0.05

k:

RES.5

100 k:

100.0

±0.5

k:

Figure 1 Testing Circuit and Test Equipment

SC202S

SUPPLY SENSOR

+

-

G

11

12

14

13

15

16

24V DC

+

100 :

DC source

+

-

-

DC

Decade

Box 1

Note 1: Cable connected to sensor input should be conductivity detector cable of

length 2.1 ±0.1 m. Connect pins 13 and 14, also 15 and 16, to Decade box 2
terminals. Use shielded cable, and connect shield to pin 14.

Note 1

Decade

Box 2

Ground