Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual
Page 95
Advertising

IM 12D08B02-01E
Appendix 3-7
3/3
QIS 12D08B02-21E
Table 3
Message display
Decade box 2 resistance
Data display
RES.1
10 :
10.00
±0.05
:
RES.2
100 :
100.0
±0.5
:
RES.3
1 k:
1.000
±0.005
k:
RES.4
10 k:
10.00
±0.05
k:
RES.5
100 k:
100.0
±0.5
k:
Figure 1 Testing Circuit and Test Equipment
SC202S
SUPPLY SENSOR
+
-
G
11
12
14
13
15
16
24V DC
+
100 :
DC source
+
-
-
DC
Decade
Box 1
Note 1: Cable connected to sensor input should be conductivity detector cable of
length 2.1 ±0.1 m. Connect pins 13 and 14, also 15 and 16, to Decade box 2
terminals. Use shielded cable, and connect shield to pin 14.
Note 1
Decade
Box 2
Ground
Milli-
ammeter
Advertising