Measuring equipment configuration commands – GW Instek GLC-9000 User Manual User Manual
Page 123

REMOTE CONTROL
123
Measuring Equipment Configuration Commands
EQUipment
Sets or queries the EUT class.
Syntax
EQUipment {CLAss1|CLAss2|INTernal}
Query Syntax
EQUipment?
Query Return
Returns the equipment class of the EUT:
CLAss1|CLAss2|INTernal
CLAss1: CLASS I
CLAss2: CLASS II
INTernal:Internally
powered
Example
EQUipment CLAss1
Sets the class of the EUT to “CLASS I”.
Query Example
EQUipment?
Return: CLASS1
Returns the current EUT class.
EQUipment:TYPE
Sets or queries the applied part of the EUT.
Note: This command can only be used with medical networks/Network B
Syntax
EQUipment:TYPE {B|BF|CF}
Query Syntax
EQUipment:TYPE?
Query Return
Returns the applied part of the EUT: B|BF|CF (Medical
Network only)
B: B Type
BF: BF Type
CF: CF Type
Example
EQUipment:TYPE BF
The medical network is set to type BF.
Query Example
EQUipment:TYPE?
Return: BF
Type BF is the currently applied part of the EUT for
network B/Medical networks.