Measurement Mode Command
MODE
Set or queries the measurement mode.
Note: Different measuring networks have different measurement modes.
Syntax
MODE
EARTh|ENCLosure1(ENCL1)|ENCLosure2(ENCL2)
|ENCLosure3(ENCL3)|PATient1(PAT1)|PATient2(PAT2)
|PATient3(PAT3)|PAUXiliary(PAUX)
Query Syntax
MODE?
Query Return
Returns the measurement mode:
EARTH|ENCLOSURE1|ENCLOSURE2|ENCLOSURE3|P
ATIENT1|PATIENT2|PATIENT3|PAUXILIARY
When measurement network A,B,C,D,E,G,H or I is
selected:
EARTh
Earth leakage current.
ENCLosure1
Enclosure to earth leakage current.
ENCLosure2
Enclosure to enclosure leakage
current.
ENCLosure3
Enclosure to line leakage current.
When measurement network F is selected:
EARTh
Earth leakage current
ENCLosure1
Enclosure to earth leakage current
ENCLosure2
Enclosure to enclosure leakage
current
PATient1
Patient leakage current I
PATient2
Patient leakage current II
PATient3
Patient leakage current III
PAUXiliary
Patient Auxiliary Current
Example
MODE EARTH
Sets the measurement mode to Earth leakage current.