GW Instek GLC-9000 User Manual User Manual
Page 148

GLC-9000 User Manual
148
<Num value9>
Auto Measurement test items,
range 1~512:
256
128
64
32
16
8
4
2
1
Bit8
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
NLINe
LLINe
Neg
phase
Pos
phase
RAPPly NAPPly EARTh
POWerso
urce
NORMal
<Num value10> Auto measurement measuring time
in seconds {NR1}
<Num value11> Auto measurement wait time in
seconds {NR1}
Note:
Measurement conditions and Power supply polarity
(<num value7&8>) are returned when a manual
measurement is set.
<num value9> ~ <num value11>are the return values
when an automatic measurement is set.
0 is returned when there is no data.
Query Example
SYSTem:FILE? 6
Return:
Panel06,NO-6,CLA2,B,F,1,0,1,2,+2.500E-02,+0.400E-
02,0,0,123,1,1
The 6th file has the following configuration:
File name
Panel06
File No.
NO-6
Equipment class
CLASS-II
Network (medical)
B
Network,(circuit)
F network
Measurement mode
Enclosure - earth
leakage
Measurement method
automatic
Leakage current type
DC
Leakage current range
10.00mA