GW Instek GLC-9000 User Manual User Manual
Page 157

REMOTE CONTROL
157
Under the following situations, the status byte register will be
cleared:
Using the *CLS command.
Reading the Standard Event Register
Under the following conditions the Status Byte Enable Register will
be cleared:
Using the *SRE 0 command.
Using the STB? Query:
If bits 2-5 of the Status Byte Register are set to 0, using the
*STB? query will clear the Request Service Bit (bit 6).
Using the *OPC command:
To see if a command has completed, the *OPC command or query
(OPC?) can be used. The OPC command will set bit “0” of the Standard
Event Register to 1 when the operation proceeding the OPC command
has completed. In contrast, the OPC query will not set bit “0”of the
Standard Event Register, instead, it will return “1”.
A query error occurs if the return value makes the output queue exceed
the output queue size. A command error occurs if data follows a
command.