Medical equipment, Earth leakage current – GW Instek GLC-9000 User Manual User Manual
Page 44

GLC-9000 User Manual
44
Medical Equipment
MD-F
CLASS I
CLASS II
Internal Power
Type B Type F* Type B
Type F* Type B Type F*
Earth leakage current
Not Used
N/A
N/A
Enclosure and
earth leakage
current
Normal
T2
T2
T2
T2
T2
T2
Fault
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
T2,T3
Enclosure and
enclosure leakage
current
Normal
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Fault
T1~3
T1~3
T1~3
T1~3
T1~3
T1~3
Patient auxiliary current
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
T1,T2
Patient leakage current I
T2
T2
T2
T2
T1,T2
T1,T2
Patient leakage current II
T2,T3
N/A
T2,T3
N/A
T2,T3
N/A
Patient leakage current III
N/A
T2
N/A
T2
N/A
T2
*Type F = Type BF and Type CF.
Earth Leakage Current
Network
N/A.
Connection
1. Connect the EUT power cord to the GLC-9000
as shown in the diagram below.
Measurement
Setup
Confirm all settings including MD, leakage current
mode, measurement time, upper and lower limits,
and other parameters.