Leakage current modes – GW Instek GLC-9000 User Manual User Manual
Page 29

OVERVIEW
29
Leakage Current Modes
Type
Normal
Condition
Single
Fault
Fault Description
Notes
Earth Leakage
Current
Yes
Yes
Power line disconnected.
Class 1
only
Enclosure to
Earth leakage
current
Yes
Yes
1. Power line
disconnected.
2. Disconnected earth
protection*.
3.110% of the rated power
supply voltage applied
to the signal
input/output parts and
earth.
Class 1
only*
Enclosure to
Enclosure
leakage current
Yes
Yes
Enclosure to
line leakage
current.
Yes
Yes
Patient Auxiliary
Current
Yes
Yes
1. Power Line
disconnected.
2. Earth disconnected*.
Class 1
only*
Patient Leakage
Current I
Yes
Yes
Patient Leakage
Current II
No
Yes
110% of the rated
power supply voltage
applied to the signal
input/output parts and
earth.
Type B
applied
part
Patient Leakage
Current III
No
Yes
110% of the rated
power supply voltage
is applied to the F-type
applied part and earth.
Type F
applied
part