Measurement items – GW Instek GLC-9000 User Manual User Manual
Page 126

GLC-9000 User Manual
126
OFF: RC filter is OFF, pure 1k resistance is enabled.
Example
CONFigure:FILTer OFF
Turns the RC filter OFF for Measuring Network F.
Query Example
CONFigure:FILTer ?
Return: OFF
Returns the RC filter status. The RC filter is turned off.
Measurement Items
CONFigure:COMParator
Sets the current measuring mode’s upper and lower limit.
Syntax
CONFigure:COMParator {[<NR3>[|,<NR3>]]}
Query Syntax
CONFigure:COMParator?
Query Return
Returns the current measuring mode upper and lower
limit. The first parameter is the upper limit, the
second is the lower limit.
<NR3>: Range: +4.000E-6 ~ +20.00E-03 (in Amps)
Example
CONFigure:COMParator +4.000E-03,+100.0E-06
Set the upper limit to 4mA and the lower limit to
100uA.
Query Example
CONFigure:COMParator?
Return: +4.000e-03,+1.000E-04
Returns an upper limit of 4mA and a lower limit of
100uA.
CONFigure:CURRent
Sets or queries the leakage current type.
Syntax
CONFigure:CURRent {ACDC|AC|DC|ACPeak}
Query Syntax
CONFigure:CURRent?