Enclosure and line leakage current – GW Instek GLC-9000 User Manual User Manual
Page 50

GLC-9000 User Manual
50
Enclosure and Line Leakage Current
Network
Non medical network / Medical network
Connection
As illustrated below, ensure that the power source
and test leads are properly connected.
Non Medical
Network
1. Connect the test lead to T2 terminal. Position
the test lead on an ungrounded section of the
enclosure.
The T2 terminal is high voltage. Avoid contact with the
terminal. The T2 terminal should not be connected
with an earth conductor.
Measurement
Setup
Confirm all settings including MD, leakage current
mode, measurement time, upper and lower limits,
and other parameters.
Panel Operation
2. Turn on the circuit breaker.
Connect the EUT power
terminal.
GLC-9000
3. Press the START button to
start measurements.
GLC-9000