GW Instek GLC-9000 User Manual User Manual
Page 135

REMOTE CONTROL
135
NLINe
Application of voltage from the
neutral line. Normal neutral line
connection.
Example
CONFigure: AMITem <101>
256
128
64
32
16
8
4
2
1
Bit8
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
0
0
1
1
0
0
1
0
1
Automatic measurement items include: normal power
supply, disconnected earth line, normal polarity, and
reverse polarity.
Query Example
CONFigure: AMITem?
Return: 4
Power source disconnected is the current
measurement setting.
CONFigure:AMTime
Set or query the auto measurement duration, ranging from 1~300
seconds.
Note: Auto measurement configuration cannot be used in manual mode.
Syntax
CONFigure:AMTime <NR1>
Query Syntax
CONFigure:AMTime?
Query Return
Returns the auto measurement time value.
<NR1>
1~300 secs.
Example
CONFigure:AMTime 2
Set the duration of the auto measurement to 2 secs.
Query Example
CONFigure:AMTime?
2
Returns the auto measurement time (2 seconds).