Measurement data command – GW Instek GLC-9000 User Manual User Manual

Page 137

Advertising
background image

REMOTE CONTROL

137

Measurement Data Command

MEASure:AUTO

Queries the maximum value after auto measurement.
Note: This command can only be used for automatic testing.

Query Syntax

MEASure:AUTO?

Query Return

Returns the file name and 5 values < value1>,
<value2>, < value3>, <value4>, < value5>.

<value1>

The maximum value in Amps

<value2>

Power supply polarity:
0: Positive polarity
1: Negative polarity

<value3>

Equipment status:
0: Normal conditions
1: Disconnected live line.
2: Disconnected earth line.
3: Normal neutral line connection
conditions, 110% voltage
application.
4: Negative phase, 110% voltage
application.
5: Normal live line connection
conditions. Application of voltage
from the live line.
6:Normal neutral line connection
conditions. Application of voltage
from the neutral line.

<value4>

Leakage current type:
0: AC
1: DC
2: AC+DC
3: ACpeak