Key Features
International
Standards and
Regulations
The GLC-9000 has nine measurement networks
(Measuring Devices: MD) supporting GB/12113,
IEC/UL and other international standards for
electrical products:
(1) MD-A:IEC60990
(2) MD-B:IEC60990
(3) MD-C:IEC60990
(4) MD-D:UL
(5) MD-E(1kΩ):-general application
(6) MD-F: Medical
(7) MD-G:UL
(8) MD-H:(2kΩ)-general application
(9) MD-I:JIS
Measurement
Modes
There are a number of leakage current
measurement tests covering general electrical
equipment and medical electrical equipment.
(1) Earth Leakage Current.
(2) Leakage current from enclosure and earth.
(3) Leakage current from enclosure to enclosure.
(4) Leakage current from enclosure to line.
(5) Patient Auxiliary Current*.
(6) Patient leakage current I*.
(7) Patient leakage current II*.
(8) Patient leakage current III*.
*Tests 5,6,7,8 are applicable to medical MD-F
devices.
Leakage Current
Types
Leakage current measurement modes:
DC, AC, AC+DC, AC Peak.