Save data command – GW Instek GLC-9000 User Manual User Manual
Page 138

GLC-9000 User Manual
138
<value5>
Judgment Value:
0: Measurement is within upper
and lower judgment limits (PASS)
1: Measurement is greater than the
upper limit (FAIL-U)
2: Measurement is less than the
lower limit (FAIL-L)
Query Example
MEASure:AUTO?
File_01
+1.031E-03,0,0,1,0,+1.024E-03,1,0,1,0
+1.040E-03,0,1,1,1,+1.019E-03,1,1,1,0
+1.013E-03,0,2,1,2,+1.027E-03,1,2,1,0
Return value examples are shown in the table below:
Maximum
Power
polarity
Equipment
Status
Leakage
current type
Judgment
1.031mA
Positive
Normal
DC
PASS
1.024mA
Negative
Normal
DC
PASS
1.040mA
Positive
Power line
disconnected
DC
FAIL-U
1.019mA
Negative
Power line
disconnected
DC
PASS
1.013mA
Positive
Earth
disconnected
DC
FAIL-L
1.027mA
Negative
Earth
disconnected
DC
PASS
Save Data Command
MEMory:NUMBer
Query the data where the file no. is recorded and saved.
Query Syntax
MEMory:NUMBer?
Query Return
Returns the number of data files, ranging from 1~100.