Manual measurement commands – GW Instek GLC-9000 User Manual User Manual
Page 129

REMOTE CONTROL
129
Manual Measurement Commands
CONFigure:CONDition
Sets or queries the EUT status when in manual measurement mode.
Note: The CONFigure:CONDition command or query can only be used in
manual measurement mode.
Syntax
CONFigure:CONDition
{NORMal|EARTh|POWersource|NAPPly|RAPPly|LLINe
|NLINe}
Query Syntax
CONFigure:CONDition?
Query Return
Returns the EUT status when in manual measurement
mode.
NORMal|EARTh|POWersource|NAPPly|RAPPly|LLINe|
NLINe
NORMal
Under normal conditions.
EARTh
Disconnected earth line.
POWersource
Disconnected live line.
NAPPly
Normal connection, positive phase,
110% voltage application.
RAPPly
Negative phase, 110% voltage
application.
LLINe
Application of voltage from the live
line. Normal live line connection.
NLINe
Application of voltage from the
neutral line. Normal neutral line
connection.
Example
CONFigure:CONDition NORMal
Set the leakage current test to normal conditions.
Query Example
CONFigure:CONDition ?
Return: NORMAL
The leakage current test is set to normal conditions.