Query Example
MODE?
Return: EARTH
Earth leakage current is the current measurement
mode.
Measurement Commands
CONFigure:AUTO
Configures or queries the current measurement mode. When the
measurement mode is set to Auto, a measurement can be started after
any of the following:
The
Start
key on the panel has been pressed.
The START line of the EXT I/O line is sent low
The STARt command is used.
Syntax
CONFigure:AUTO {ON|OFF}
Query Syntax
CONFigure:AUTO?
Query Return
Returns the current measurement mode as a string:
(ON|OFF)
ON: Automatic mode
OFF: Manual mode
Example
CONFigure:AUTO OFF
Sets the measurement mode to manual mode.
Query Example
CONFigure:AUTO?
Return: OFF
The current measurement mode is manual.
CONFigure:FILTer
Sets or queries whether Network F has the RC network enabled/disabled.
If the RC network is OFF, a 1k pure resistance filter is used.
Syntax
CONFigure:FILTer {ON|OFF}
Query Syntax
CONFigure: FILTer?
Query Return
Returns a string indicating if the RC network is ON or
OFF.
ON: RC filter is ON.