System setup command – GW Instek GLC-9000 User Manual User Manual
Page 141

REMOTE CONTROL
141
1.040mA
Positive
Power line
disconnected
DC
FAIL-U
1.019mA
Negative
Power line
disconnected
DC
PASS
1.013mA
Positive
Earth
disconnected
DC
FAIL-L
1.027mA
Negative
Earth
disconnected
DC
PASS
Note
If there’s no return value for the file no. query, 0 will be
returned.
MEMory:SAVE:AUTO
Saves auto measurement results, including file name, file no.,
instrumentation class, medical network application, network,
measurement mode, measurement type, leakage current, leakage current
range, maximum limit, minimum limit, measuring conditions, power
supply polarity and auto measurement items.
Syntax
MEMory:SAVE:AUTO
Example
MEMory:SAVE:AUTO
Saves auto measurement results.
System Setup Command
SYStem:BACKlight
Sets or queries the backlight to turn off automatically after a set time.
Syntax
SYStem:BACKlight
<NR1>
Range: 0~30
Query Syntax
SYStem:BACKlight?
Query Return
Returns the wait time (in minutes) before the
backlight will automatically turn off.
0
Infinite (backlight will not
automatically turn off).
1~30
Time in minutes before the
backlight will turn off.