Specifications, Appendix, And ½ respectively. dc – GW Instek GLC-9000 User Manual User Manual
Page 177: Ac/ ac+dc, Ac peak

APPENDIX
177
Specifications
Specification accuracy is only applicable when the GLC-9000 has
been warmed up for 30 minutes and has an operating temperature
of +18°C – +28°C. The specifications below are based on a 1kΩ
purely resistive network. For networks G and H scale the range by
1/1.5
and ½ respectively.
DC
Ranges
Range
Resolution
Accuracy
25.00mA
5.00mA~25.00mA
10uA
±(0.2%rdg+3dgt)
5.000mA
0.500mA~5.000mA
1uA
±(0.2%rdg+3dgt)
500.0uA
50.0uA~500.0uA
0.1uA
±1.0%fs
50.00uA
4.00uA~50.00uA
0.01uA
±1.0%fs
AC/ AC+DC
Ranges
Range
Resolution
Accuracy
10Hz≦f≦100kHz 100kHz<f≦1MHz
25.00mA
5.00mA~25.00mA
10uA
±(2.0%rdg+6dgt)
±(2.0%rdg+10dgt)
5.000mA
0.500mA~5.000mA
1uA
±(2.0%rdg+6dgt)
±(2.0%rdg+10dgt)
500.0uA
50.0uA~500.0uA
0.1uA
±(2.0%rdg+6dgt)
±(2.0%rdg+10dgt)
50.00uA
4.00uA~50.00uA
0.01uA
±2.0%fs
±2.0%fs
AC Peak
Ranges
Range
Resolution
Accuracy
20Hz≦f≦1kHz
1kHz<f≦10kHz
75.0mA
10.0mA~75.0mA
100uA
±(2.0%rdg+2dgt)
±(5.0%rdg+10dgt)
10.00mA
1.00mA~10.00mA
10uA
±(2.0%rdg+2dgt)
±(5.0%rdg+10dgt)
1.000mA
0.500mA~1.000mA
1uA
±2.5%fs
±5.0%fs
500.0uA
40.00uA~500.0uA
0.1uA
±4.0%fs
±5.0%fs
EUT Voltage / Current
Ranges
Range
Resolution
Accuracy
300V
85V~300V
0.1V
±(2%rdg+10dgt)
10A
0.5A~10A
0.1A
±(2%rdg+5dgt)
Voltage
Ranges
Range
Resolution
Accuracy
AC
10.0~300.0V
0.1V
±(3%rdg+2V)
DC
10.0~300.0V
0.1V
±(3%rdg+2V)
AC+DC
10.0~300.0V
0.1V
±(3%rdg+2V)
AC Peck
15.0~430.0V
0.1V
±(3%rdg+2V)