GW Instek GLC-9000 User Manual User Manual
Page 74

GLC-9000 User Manual
74
Note
The leakage current range depends on the measuring
network used. The table below shows the maximum
and minimum values of each range for each network
type.
MD A~F,I
AC, DC,
AC+DC
Range
25.00mA
5.000mA
500.0uA
50.00uA
Maximum 25.00mA 5.000mA 500.0uA
50.00uA
Minimum 4.00mA
0.400mA
40.00uA
4.00uA
AC Peak
Range
75.0mA
10.00mA
1.000mA
500.0uA
Maximum 75.0mA
10.00mA
1.000mA
500.0uA
Minimum 8.0mA
0.80mA
0.100mA
40.0uA
MD G
AC, DC,
AC+DC
Range
25.00mA
5.000mA
500.0uA
50.00uA
Maximum 16.00mA 3.300mA 330.0uA
33.00uA
Minimum 3.00mA
0.300mA
30.00uA
4.00uA
AC Peak
Range
75.0mA
10.00mA
1.000mA
500.0uA
Maximum 50.0mA
6.60mA
0.660mA
330.0uA
Minimum 6.0mA
0.60mA
0.070mA
30.0uA
MD H
AC, DC,
AC+DC
Range
25.00mA
5.000mA
500.0uA
50.00uA
Maximum 12.50mA 2.500mA 250.0uA
25.00uA
Minimum 2.00mA
0.200mA
20.0uA
4.00uA
AC Peak
Range
75.0mA
10.00mA
1.000mA
500.0uA
Maximum 37.5mA
5.00mA
0.500mA
250.0uA
Minimum 4.0mA
0.40mA
0.050mA
20.0uA