6 .8 .3 solartron/sarasota transmitter – KROHNE Summit 8800 Vol 2 User Manual
Page 105
105
www.krohne.com
08/2013 - MA SUMMIT 8800 Vol2 R02 en
SUMMIT 8800
STREAM CONFIGURATION
06
Figure 99 Viscosity
6 .8 .3 Solartron/Sarasota transmitter
For each Solartron/Sarasota transmitter selected, an additional icon “ρtp sensor” will appear in
the left hand list:
Liquid
Gas
Steam
Liquid
Gas
Steam
The user will be required to enter the transducers calibration constants:
Figure 100 Density transducer parameters
For details on these parameters, please refer to the transducer calibration and instruction
manual.
The pressure and temperature at the density point must be also specified. Their selection is
almost identical to the stream pressure and temperature selection. Please note that in this case
only 2 pressure and temperature sensors are possible but that also the pressure and tempera-
ture at the meter can be used.