Figure 28 Configure HART inputs
Select the HART loop required as loop 1 or 2. In “options” select in Master 1 or 2 or Burst mode.
Please note that in burst mode only 1 transmitter can be chosen, the other transmitters are
greyed out..
Set the number of retries and the timeout (in seconds) before an error must be generated.
Then for each transmitter set under “Configure Transmitter”:
• Assign a short id address: 0 when in burst mode, between 1 and 15 otherwise. Make sure the
actual transmitters are programmed accordingly.
• Select the primary variable to be read from this transmitter from pull down tree.
• Select the other variables if the transmitter is a multi-variable type.
5 .1 .2 Analog Inputs
Only for the analog input board there are 4 analog inputs. For details, see volume 1.
Configure each sensor input as follows:
Figure 29 Configure analog input