SUMMIT 8800

PROVER

10

Used switches

Sequence of switches used or the average of all detectors

Switches volume

Prover base volume at switch point

K-factor calculation

Method of K-factor calculation selectable

Confidence

User confidence level of calculation implemented

Required runs

Least amount of run required for a sucessful prove

Pulse interpolation

Method for pulse interpolation – non or double chronometry

Minimum pulses

Minimum allowable pulses for a prove, default 10,000

Max KF deviation

Maximum deviation limit for new K-factor compare to previous

value

Max MF deviation

Maximum deviation limit for new MF compare to previous value

Please note that some parameters will not be needed, depending on previous choices.

The K-factor calculation method can be selected from

• Traditional repeatability

• Statistical repeatability

• Uncertainty calculations.

Provers are normally set to traditional repeatability. Uncertainty can also be selected but this

requires a high repeatability and more runs than traditional repeatability.

Pipe correction

This section is for liquid provers only,

Here the user must enter the corrections for the liquid prover stream for CTSp (prover expan-

sion due to temperature) and CPSp (prover metal expansion due to pressure)

Figure 152 Prover calculations, pipe correction