6 .2 .2 .1 meter input – KROHNE Summit 8800 Vol 2 User Manual
Page 76

76
www.krohne.com
08/2013 - MA SUMMIT 8800 Vol2 R02 en
SUMMIT 8800
STREAM CONFIGURATION
06
Meter input
the meter connected and its basic settings
Pulse input
the preference for serial or pulse input and the pulse input setting if used
Meter correction
the correction for expansion of the meter body for pressure and
temperature
Meter information
the identification of the meter
Most smart UFM have the option to send the flow data via a serial link and via pulses. In case
pulses are used, then the API level has to be selected. For liquid the pulse input can either be
API 5.5 level A or level B to E, for gas and steam API 5.5 Level B to E only.
API 5.5 Level B to E
for single or dual pulse with the same of different frequencies and pulse
monitoring.
API 5.5 Level A
for dual pulse with pulse correction.
See also chapter 3.1.
6 .2 .2 .1 Meter input
Define which meter type or manufacturer are applicable along with the associated parameters
associated with the meter.
Figure 68 Example ultrasonic meter input section