3 .11 sampling – KROHNE Summit 8800 Vol 2 User Manual

Page 46

Advertising
background image

46

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol2 R02 en

SUMMIT 8800

METERING PRINCIPLES

03

Description

C01

Start Run

C02

Rotate 4 way valve forward

C03

Are valves correctly positioned?

If not abort run (jump to C21)

C04

Detector switch 1 triggered

C05

Start counting pulses

C06

Detector switch 2 triggered

C07

Stop counting pulses

C08

Wait for ball to reach chamber

C09

Run timer timeout activate?

If so abort run (jump to C21)

C10

Rotate 4 way valve reverse

C11

Valves correctly positioned?

C12

Detector switch 2 triggered

If not abort run (jump to C21)

C13

Start counting pulses

C14

Detector switch 1 triggered

C15

Stop counting pulses

C16

Wait for ball to reach chamber

C17

Calculate pulses

C18

Release all flow control valves

C19

Enough pulses received?

If not abort run (jump to C22)

C20

Run completed successfully

C21

Release all flow control valves

C22

Run completed un-successfully

3 .11 Sampling

Fixed time period

Activated during defined period with on and off timings

Duration

Actived for a fixed lenght of time

Batch

Quantity measurement based on parcel size

Continuous

Continuously running and obtaining samples