SUMMIT 8800
STREAM CONFIGURATION
06
Figure 106 Z-table
The contents of the table can be imported from another configuration by clicking on the button
“Import” and selecting the other configuration.
6 .9 Liquid line density at the metering conditions
Liquid
Gas
Steam
Typically the line density will be measured and from this the meter density will be calculated.
This will be done in two steps: from line to base density and from base to meter density, both
correcting for pressure and temperature.
For the calculation of the line density at metering conditions, there are three options for both
pressure and temperature correction:
Keypad
The user defined value for this range will be choosen as given.
Norsok I-105 rev2
The Norsok calculation will be used
API chapter 12.2.5.1&2
The API calculation will be used
In case of keypad the user defined value has to be filled-in:
or
Figure 107 Meter line density, keypad
All other parameter are not needed and will be greyed-out